Conductivity Sensor Technical Guide: What Is Water Conductivity, Testing & Calibration Instructions.
What is conductivity?
Conductivity is the ability of water to conduct electricity. This conduction requires ions in the water to be charge carriers. Pure water contains no ions (aside from the occasional rogue water molecule), and so is a very poor conductor of electricity. But as water picks up chemical salts, those salts add ions to the water when they dissolve. So as the ionic strength of the water rises, so does conductivity – the “saltier” the water, the higher the conductivity.
Note that a conductivity measurement does not necessarily tell you the concentration of ions in the water, because different ions have different contributions to conductivity. Large, ungainly ions, like acetate, do not help conduct electricity well. Small, fast ions, like H+, make a much higher per-ion contribution to conductivity. Since natural waters are usually ever-changing mixtures of ions, a conductivity reading is only a relative measurement of ion concentration.
Salinity is often calculated from conductivity by assuming that the water is very similar in chemical constituency to sea water. In sea water, a conductivity of 55,000 µS/cm is roughly the same thing as a salinity of 35 PSS (PSS is the practical salinity scale, which has replaced the traditional ppt salinity units). Salinity is technically defined only for small dilutions and concentrations of salt water, and so is seldom used in fresh-water work.
Why is conductivity important in water quality monitoring?
From a chemical standpoint, conductivity readings can be used to monitor the stability of a water body. An unchanging conductivity usually means that no chemicals are being added to, or subtracted from, the water. A spike in conductivity can indicate an increase in ions from a pollution source, such as a waste-water outfall, agricultural runoff, or tidal intrusion. A dip in conductivity can indicate a reduction of chemical inputs and/or an increase in fresh-water input (from a rainfall event, for instance).
Conductivity readings can also be used to judge biological health because some plant and animal species are sensitive to high, or low, conductivities. Oysters, for instance, do not do well in significantly diluted sea water. Fresh-water game fish do not do well in salty waters.
Conductivity readings can also be used inferentially. For instance, it is common in monitoring fresh-water beaches to assume that a sudden increase in conductivity is due to waste-water contamination or rainfall run-off – both of which are usually associated with high bacteria counts. Thus, the conductivity spike may be reason to close the beaches.
Finally, conductivity readings are used to correct water-level measurements and dissolved-oxygen readings. Increasing conductivity means increasing water density, so water-level measurements must be decreased to compensate. Increasing conductivity also reduces the solubility of oxygen in water, so calculations of percent-saturation of oxygen must be increased to compensate.
How to Measure Water Conductivity
Two mainstream sensor types are used for conductivity detection: toroidal sensors and four-electrode sensors. Toroidal sensors are rarely integrated into multi-parameter water monitors due to their low sensitivity under low-conductivity freshwater conditions. Four-electrode sensors calculate conductivity by measuring the current needed to maintain a fixed voltage between two pairs of electrodes with fixed spacing. Both sensor types deliver reliable test results for most natural water monitoring projects.
How do you calibrate conductivity sensors?
Conductivity sensors are calibrated with one standard, and that standard is almost always a known concentration of potassium chloride (KCl). The value at which calibration is made depends on the application – for very fresh waters, a low conductivity standard is best, and vice-versa. Choose a calibration standard that is a bit higher than the highest reading you expect to see in the field.
Key Factors Affecting Conductivity Test Results
Conductivity values are highly temperature-dependent: identical water samples exhibit lower conductivity at 15 °C than at 30 °C. To enable horizontal comparison of test data across different sampling sites and times, raw conductivity readings are standardized to a reference temperature of 25 °C via temperature compensation algorithms. Temperature-corrected conductivity data calibrated to 25 °C is defined as specific conductance, distinguishing it from unadjusted raw conductivity measurements.